| + | ====== Calibrating a Digilent Test and Measurement Device ====== | ||
| + | ~~TechArticle~~ | ||
| + | |||
| + | While all Digilent Test and Measurement devices are factory calibrated, they can be recalibrated so that you can ensure that you are making the most accurate measurements possible. | ||
| + | |||
| + | ---- | ||
| + | ==== Inventory ==== | ||
| + | * A Digilent Test and Measurement device to be calibrated. This guide supports the following devices: | ||
| + | * [[test-and-measurement: | ||
| + | * [[test-and-measurement: | ||
| + | * [[test-and-measurement: | ||
| + | * //Under Construction//: | ||
| + | * A Mac, Windows, or Linux computer with [[software: | ||
| + | * If WaveForms is not yet installed, please review your device' | ||
| + | * A reliable [[https:// | ||
| + | * DMM probes with grabber ends | ||
| + | * A [[https:// | ||
| + | * [[https:// | ||
| + | |||
| + | ---- | ||
| + | **Note:** //While all Digilent Test and Measurement devices are factory calibrated and can be calibrated by the user in WaveForms' | ||
| + | ---- | ||
| + | ====== Guide ====== | ||
| + | ==== 1. Device Manager ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | In the **Settings** menu in the bar at the top of the WaveForms window, click on **Device Manager** to open the device manager window. | ||
| + | |||
| + | Once the Device Manager window opens, you can explore all of the features of the device manager. | ||
| + | |||
| + | In the top half of the window, the table contains the devices that you can use with WaveForms. If you have a device plugged in, it will appear here. | ||
| + | |||
| + | Plug in your device so that it shows up in the device manager. If it is the only device connected, it will be the only entry in the table that has a serial number. | ||
| + | </ | ||
| + | {{ reference: | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 2. Rename Tool ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | The first tool to explore is the rename tool. In the device manager, click on " | ||
| + | </ | ||
| + | {{ reference: | ||
| + | </ | ||
| + | |||
| + | <WRAP group>< | ||
| + | For example, we named ours " | ||
| + | |||
| + | The Rename tool can be useful if you have more than one device or are collaborating in a large group where there will be multiple devices present. | ||
| + | </ | ||
| + | {{ reference: | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 3. Calibration Window ==== | ||
| + | |||
| + | //Next, we are going to explore the calibration feature to ensure the best possible accuracy for captured data.// | ||
| + | |||
| + | <WRAP group>< | ||
| + | In the Device Manager window, click " | ||
| + | |||
| + | The first tab you will see in the calibration window is the " | ||
| + | |||
| + | There is also the " | ||
| + | |||
| + | The Parameters and References tabs can also be seen. These are the values that define the existing calibration. It is highly recommended that you do **NOT** manually change these values. Instead, we will do a full calibration using the Calibration Wizard. | ||
| + | |||
| + | **Note:** //The list of calibrations that can be performed, as seen in the screenshot to the right, is not representative of all devices.// | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 4. Device Calibration Wizard ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | To open the Device Calibration Wizard, click " | ||
| + | |||
| + | When the wizard opens, the start screen appears, as in the figure to the right, which tells you what you will need in order to calibrate your device. Click **Next**. | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | |||
| + | ==== 5. Waveform Generator Linearity ==== | ||
| + | <WRAP group> <WRAP column half> | ||
| + | **Note:** //This calibration is available only on some devices. If does not appear in the calibration wizard, skip ahead to [[# | ||
| + | |||
| + | Available on some devices, this calibration occurs entirely within the device itself, and no external connections or measurements need to be made. Click **Next** to perform it and continue. | ||
| + | </ | ||
| + | {{ reference: | ||
| + | </ | ||
| + | ---- | ||
| + | ==== 6. Calibrate Waveform Generator 1 ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | The first thing you'll need to do calibrate is the Wavegen Channel 1 at DC. To do this, connect the Wavegen Channel 1 pin to your digital multimeter. | ||
| + | |||
| + | Connect the device' | ||
| + | |||
| + | Connect the negative side of the multimeter to the device' | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | <WRAP group>< | ||
| + | Read the value on the multimeter. As indicated by the Calibration Wizard, it should be around 0 V. | ||
| + | |||
| + | Enter the value you read off of the multimeter into the Calibration Wizard, and click **Next**. | ||
| + | |||
| + | Repeat this process until the Calibration Wizard indicates to move on to waveform generator 2 (10 total measurements). | ||
| + | |||
| + | **Note:** //Default values in the " | ||
| + | |||
| + | **Note:** //The Analog Discovery Pro (ADP3450/ | ||
| + | |||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 7. Calibrate Waveform Generator 2 ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | Create the same circuit as before, using waveform generator 2 instead of waveform generator 1. | ||
| + | |||
| + | To do this, swap out the waveform generator analog output 1 pin with the waveform generator analog output 2 pin. Keep the existing connection between the DMM and the device ground. If using a device with MTE cables, the new connection will be to the yellow wire with white stripes. | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | <WRAP group>< | ||
| + | Read the value on the DMM, it should be around 0 V. Enter this value into the Calibration Wizard and hit **Next**. | ||
| + | |||
| + | Repeat this procedure until the Calibration Wizard indicates to move on to Oscilloscope Calibration (10 total measurements). | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 8. Oscilloscope ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | The next step in the calibration process is to zero the oscilloscope. | ||
| + | |||
| + | As indicated by the Calibration window, connect each of the oscilloscope channels to ground. All channels will be calibrated at the same time, meaning that (if the device uses BNC cables) you will need a BNC oscilloscope probe for each channel. | ||
| + | |||
| + | For example, if using a two-channel device with MTE cables, connect the positive end of Channel 1 (orange cable), positive end of Channel 2 (blue cable), negative end of Channel 1 (orange cable with white stripes), negative end of Channel 2 (blue cable with white stripes), and the ground (black cable). | ||
| + | |||
| + | If using BNC cables, set the attenuation for each BNC probe to 1X. | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | <WRAP group>< | ||
| + | Click **Next** to start the calibration. | ||
| + | |||
| + | The oscilloscope zero calibration is automatic, so all you have to do is wait for it to finish (this may take several minutes). A progress bar indicates how much of this automatic process has been completed. Once it completes, press **Next** to continue to the next step. | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | <WRAP group>< | ||
| + | Next, the oscilloscope will be calibrated, by comparing automatically-taken readings to known voltages output by the waveform generator. The first step is to set up the circuit. | ||
| + | |||
| + | Connect the positive ends of each of the oscilloscope channels to Wavegen Channel 1, and (if the channels are differential) the negative ends of the two channels of the oscilloscope to ground. | ||
| + | |||
| + | MTE cables for analog inputs are colored blue and orange and the negative ends have white stripes. | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | <WRAP group>< | ||
| + | Click **Next** to start the oscilloscope calibration. This is another automatic calibration, | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 9. Digital Supply ==== | ||
| + | **Note:** //If your device does not have a digital supply, skip ahead to [[# | ||
| + | |||
| + | <WRAP group> <WRAP column half> | ||
| + | Calibration of the digital supply requires the use of a trusted DMM to measure the voltage between the supply pin and ground. | ||
| + | </ | ||
| + | {{ reference: | ||
| + | </ | ||
| + | ---- | ||
| + | Measurements should be taken as the wizard applies a 1.2 V and 3.2 V signal to the supply pins. Enter these measurements into the text fields, and click **Next** to save them and continue. | ||
| + | <WRAP group> <WRAP column half> | ||
| + | {{ reference: | ||
| + | </ | ||
| + | {{ reference: | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 10. Positive Variable Supply ==== | ||
| + | **Note:** //If your device does not have positive and negative variable supplies, skip ahead to [[# | ||
| + | |||
| + | <WRAP group>< | ||
| + | The next things to calibrate are the variable power supplies, starting with the positive one. | ||
| + | |||
| + | Connect the positive supply (red cable) to the positive probe of the DMM. Connect a ground pin (black cable) to the negative probe of the DMM. | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | <WRAP group>< | ||
| + | Read the value off of the DMM and enter it into the text field. As indicated by the Calibration Wizard, it should be around 1. Click **Next**. | ||
| + | |||
| + | Continue taking and entering voltage readings until the Calibration Wizard indicates to move on to the negative power supply (two total measurements). | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 11. Negative Variable Supply ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | Disconnect the positive supply (red cable) and replace it with the negative supply (white cable). | ||
| + | |||
| + | Repeat the same process used on the positive supply until the calibration window indicates that the negative supply has been calibrated (two total measurements). | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 12. Frequency Compensation ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | **Note:** //This calibration is available only on some devices. If does not appear in the calibration wizard, skip ahead to [[# | ||
| + | |||
| + | Frequency compensation occurs wholly automatically. As stated in the wizard, disconnect or ground all scope inputs, and press **Next** to run the calibration procedure. | ||
| + | </ | ||
| + | {{ reference: | ||
| + | </ | ||
| + | |||
| + | <WRAP group>< | ||
| + | When the calibration is complete, click the **Exit** button to exit the calibration wizard. | ||
| + | </ | ||
| + | {{ reference: | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 13. Done Calibrating! ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | To complete the calibration, | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 14. Apply the Changes ==== | ||
| + | |||
| + | <WRAP group>< | ||
| + | In the Device Calibration window, click **Apply**, then **Yes** in the popup, to confirm the changes. | ||
| + | </ | ||
| + | {{ : | ||
| + | </ | ||
| + | |||
| + | ---- | ||
| + | ==== 15. Compare Accuracy ==== | ||
| + | |||
| + | We've hooked up the device so that Wavegen Channel 1, configured for 0 V DC, is connected to Oscilloscope Channel 1. | ||
| + | |||
| + | In the images below, you can see captures of the 0 V value with and without calibration. You can see that with calibration, | ||
| + | <columns 100% 50% 50%> | ||
| + | {{ reference: | ||
| + | < | ||
| + | < | ||
| + | {{ reference: | ||
| + | < | ||
| + | </ | ||
| + | |||
| + | |||
| + | ---- | ||
| + | ==== 16. If Necessary, Reset Calibration ==== | ||
| + | |||
| + | If you are ever dissatisfied with the calibration of your device, you can reset the calibration to factory settings, or reset all values to zero by using the Reset menu in the Device Calibration window. | ||
| + | |||
| + | ---- | ||
| + | ==== 17. Use Your Digilent Test and Measurement Device ==== | ||
| + | |||
| + | Now that you've gone through the calibration process, you can begin taking advantage of the many features of your Test and Measurement device. | ||
| + | |||
| + | {{ : | ||
| + | |||
| + | For more guide, projects, and curriculum using your Test and Measurement device, check out its Resource Center, linked from this site's [[test-and-measurement: | ||