| ====== OpenLogger - Calibration ====== | ====== OpenLogger - Calibration ====== | ||
| + | |||
| + | ===== First-Time Calibration ===== | ||
| Before being used for the first time, a new OpenLogger must first be calibrated. This process determines constants used to accurately calculate analog voltages for data points being logged, and stores these constants in non-volatile memory. | Before being used for the first time, a new OpenLogger must first be calibrated. This process determines constants used to accurately calculate analog voltages for data points being logged, and stores these constants in non-volatile memory. | ||
| - | ===== How to Calibrate | + | **Note:** //After first-time calibration, |
| + | |||
| + | ---- | ||
| <WRAP group> | <WRAP group> | ||
| Line 10: | Line 14: | ||
| When the device is connected to WaveForms Live for the first time, a prompt to calibrate the OpenLogger will appear. Click **OK**. | When the device is connected to WaveForms Live for the first time, a prompt to calibrate the OpenLogger will appear. Click **OK**. | ||
| - | **Note:** //After first-time calibration, | ||
| </ | </ | ||
| <WRAP column half> | <WRAP column half> | ||
| - | {{ :reference:instrumentation: | + | {{ reference:test-and-measurement: |
| </ | </ | ||
| </ | </ | ||
| Line 20: | Line 23: | ||
| <WRAP column half> | <WRAP column half> | ||
| === 2. === | === 2. === | ||
| - | Connect the provided 2x18 flywires to the OpenLogger with the nub side facing towards the center | + | Connect the provided 2x18 flywires to the OpenLogger with the **__nub |
| + | ---- | ||
| + | === 3. === | ||
| + | Connect the red wire on pin 6 (DC OUTPUT 1) to the solid orange wire on pin 35 (Analog Input 1). Take a look at the [[test-and-measurement: | ||
| </ | </ | ||
| <WRAP column half> | <WRAP column half> | ||
| - | {{ :reference:instrumentation:openlogger: | + | {{ /learn/instrumentation/tutorials/openlogger/calib-1.png?500 |Connect the red and orange wires}} |
| - | </ | + | |
| - | </ | + | |
| - | <WRAP group> | ||
| - | <WRAP column half> | ||
| - | === 3. === | ||
| - | Connect the red wire on pin 6 (DC OUTPUT 1) to the solid orange wire on pin 35 (Analog Input 1). Take a look at the [[: | ||
| - | </ | ||
| - | <WRAP column half> | ||
| - | {{ : | ||
| </ | </ | ||
| </ | </ | ||
| Line 43: | Line 40: | ||
| </ | </ | ||
| <WRAP column half> | <WRAP column half> | ||
| - | {{ :reference:instrumentation: | + | {{ reference:test-and-measurement: |
| </ | </ | ||
| </ | </ | ||
| Line 49: | Line 46: | ||
| <WRAP group> | <WRAP group> | ||
| <WRAP column half> | <WRAP column half> | ||
| - | === 4. === | + | === 5. === |
| Wait for the calibration initialization to complete. | Wait for the calibration initialization to complete. | ||
| </ | </ | ||
| <WRAP column half> | <WRAP column half> | ||
| - | {{ :reference:instrumentation: | + | {{ reference:test-and-measurement: |
| </ | </ | ||
| </ | </ | ||
| Line 59: | Line 56: | ||
| <WRAP group> | <WRAP group> | ||
| <WRAP column half> | <WRAP column half> | ||
| - | === 5. === | + | === 6. === |
| Connect the solid blue pin 33 (Analog Input 2 to the same red wire previously used and click the **Next** button. | Connect the solid blue pin 33 (Analog Input 2 to the same red wire previously used and click the **Next** button. | ||
| - | </ | ||
| - | <WRAP column half> | ||
| - | {{ : | ||
| - | </ | ||
| - | </ | ||
| - | <WRAP group> | ||
| - | <WRAP column half> | ||
| - | === 6. === | ||
| - | Continue through the remaining 6 analog inputs (the pink, green, gray, purple, yellow, and brown wires) until calibration is complete. Once all analog inputs have been calibrated, click the **DONE** button. | ||
| - | </ | ||
| - | </ | ||
| - | |||
| - | <WRAP group> | ||
| - | <WRAP column half> | ||
| === 7. === | === 7. === | ||
| - | A final calibration complete | + | After calibrating Analog Input 2, continue through the remaining 6 analog inputs (the pink, green, gray, purple, yellow, and brown wires) until calibration |
| </ | </ | ||
| <WRAP column half> | <WRAP column half> | ||
| - | {{ :reference:instrumentation:openlogger: | + | {{ /learn/instrumentation/tutorials/openlogger/calib-2.png?500 |Start step 2}} |
| </ | </ | ||
| </ | </ | ||
| Line 87: | Line 70: | ||
| <WRAP column half> | <WRAP column half> | ||
| === 8. === | === 8. === | ||
| + | A final calibration complete screen will appear. Click the **DONE** button - see the image to the right. | ||
| + | ---- | ||
| + | === 9. === | ||
| Click the **DONE** button at the bottom of the page to back out to the configuration screen. | Click the **DONE** button at the bottom of the page to back out to the configuration screen. | ||
| + | ---- | ||
| + | === 10. === | ||
| + | In the device configuration screen, click the back arrow at the top left of the page to return to the device selection screen. | ||
| </ | </ | ||
| <WRAP column half> | <WRAP column half> | ||
| - | {{ :reference:instrumentation: | + | {{ reference:test-and-measurement: |
| </ | </ | ||
| </ | </ | ||
| - | <WRAP group> | + | ---- |
| - | <WRAP column half> | + | ===== More Information ===== |
| - | === 9. === | + | |
| - | In the device configuration screen, click the back arrow at the top left of the page to return to the device selection screen. | + | |
| - | </ | + | |
| - | </ | + | |
| - | ---- | + | Check out the OpenLogger' |