 ====== OpenLogger - Calibration ====== ====== OpenLogger - Calibration ======
 +
 +===== First-Time Calibration =====
  
 Before being used for the first time, a new OpenLogger must first be calibrated. This process determines constants used to accurately calculate analog voltages for data points being logged, and stores these constants in non-volatile memory. Before being used for the first time, a new OpenLogger must first be calibrated. This process determines constants used to accurately calculate analog voltages for data points being logged, and stores these constants in non-volatile memory.
  
-===== How to Calibrate the OpenLogger =====+**Note:** //After first-time calibration, described in this guide, a device can be recalibrated, if desired, through its Configuration Menu, found through the Device Manager in WaveForms Live.// 
 + 
 +----
  
 <WRAP group> <WRAP group>
Line 10: Line 14:
 When the device is connected to WaveForms Live for the first time, a prompt to calibrate the OpenLogger will appear. Click **OK**. When the device is connected to WaveForms Live for the first time, a prompt to calibrate the OpenLogger will appear. Click **OK**.
  
-**Note:** //After first-time calibration, a device can be recalibrated, if desired, by clicking the CALIBRATE button in its configuration screen.// 
 </WRAP> </WRAP>
 <WRAP column half> <WRAP column half>
-{{ :reference:instrumentation:openlogger:Uncalibrated_device_prompt.png?500 |Calibrate your device}}+{{ reference:test-and-measurement:openlogger:uncalibrated_device_prompt.png?500 |Calibrate your device}}
 </WRAP> </WRAP>
 </WRAP> </WRAP>
Line 20: Line 23:
 <WRAP column half> <WRAP column half>
 === 2. === === 2. ===
-Connect the provided 2x18 flywires to the OpenLogger with the nub side facing towards the center of the OpenLogger.+Connect the provided 2x18 flywires to the OpenLogger with the **__nub side facing towards the center__** of the OpenLogger. 
 +---- 
 +=== 3. === 
 +Connect the red wire on pin 6 (DC OUTPUT 1) to the solid orange wire on pin 35 (Analog Input 1). Take a look at the [[test-and-measurement:openlogger:reference-manual#openlogger_pinout|OpenLogger Pinout]] in the Reference Manual for details on which wire corresponds to which OpenLogger pin.
 </WRAP> </WRAP>
 <WRAP column half> <WRAP column half>
-{{ :reference:instrumentation:openlogger:nub_up_to_connect_wires_correctly.jpg?500 |Connect the flywires}} +{{ /learn/instrumentation/tutorials/openlogger/calib-1.png?500 |Connect the red and orange wires}}
-</WRAP> +
-</WRAP>+
  
-<WRAP group> 
-<WRAP column half> 
-=== 3. === 
-Connect the red wire on pin 6 (DC OUTPUT 1) to the solid orange wire on pin 35 (Analog Input 1). Take a look at the [[:reference:instrumentation:openlogger:reference-manual#openlogger_pinout|OpenLogger Pinout]] in the Reference Manual for details on which wire corresponds to which OpenLogger pin. 
-</WRAP> 
-<WRAP column half> 
-{{ :reference:instrumentation:openlogger:connect_DA1_to_AN1.jpg?500 |Connect the red and orange wires}} 
 </WRAP> </WRAP>
 </WRAP> </WRAP>
Line 43: Line 40:
 </WRAP> </WRAP>
 <WRAP column half> <WRAP column half>
-{{ :reference:instrumentation:openlogger:Calibration_step_1.png?500 |Begin the calibration process}}+{{ reference:test-and-measurement:openlogger:calibration_step_1.png?500 |Begin the calibration process}}
 </WRAP> </WRAP>
 </WRAP> </WRAP>
Line 49: Line 46:
 <WRAP group> <WRAP group>
 <WRAP column half> <WRAP column half>
-=== 4. ===+=== 5. ===
 Wait for the calibration initialization to complete. Wait for the calibration initialization to complete.
 </WRAP> </WRAP>
 <WRAP column half> <WRAP column half>
-{{ :reference:instrumentation:openlogger:starting_calibration_process.png?500 |Calibration progress bar}}+{{ reference:test-and-measurement:openlogger:starting_calibration_process.png?500 |Calibration progress bar}}
 </WRAP> </WRAP>
 </WRAP> </WRAP>
Line 59: Line 56:
 <WRAP group> <WRAP group>
 <WRAP column half> <WRAP column half>
-=== 5. ===+=== 6. ===
 Connect the solid blue pin 33 (Analog Input 2 to the same red wire previously used and click the **Next** button. Connect the solid blue pin 33 (Analog Input 2 to the same red wire previously used and click the **Next** button.
-</WRAP> 
-<WRAP column half> 
-{{ :reference:instrumentation:openlogger:calibration_step_2.png?500 |Start step 2}} 
-</WRAP> 
-</WRAP> 
  
-<WRAP group> 
-<WRAP column half> 
-=== 6. === 
-Continue through the remaining 6 analog inputs (the pink, green, gray, purple, yellow, and brown wires) until calibration is complete. Once all analog inputs have been calibrated, click the **DONE** button. 
-</WRAP> 
-</WRAP> 
- 
-<WRAP group> 
-<WRAP column half> 
 === 7. === === 7. ===
-A final calibration complete screen will appear. Click the **DONE** button.+After calibrating Analog Input 2, continue through the remaining 6 analog inputs (the pink, green, gray, purple, yellow, and brown wires) until calibration is complete. Once all analog inputs have been calibrated, click the **DONE** button.
 </WRAP> </WRAP>
 <WRAP column half> <WRAP column half>
-{{ :reference:instrumentation:openlogger:calibration_complete.png?500 |Calibration complete}}+{{ /learn/instrumentation/tutorials/openlogger/calib-2.png?500 |Start step 2}}
 </WRAP> </WRAP>
 </WRAP> </WRAP>
Line 87: Line 70:
 <WRAP column half> <WRAP column half>
 === 8. === === 8. ===
 +A final calibration complete screen will appear. Click the **DONE** button - see the image to the right.
 +----
 +=== 9. ===
 Click the **DONE** button at the bottom of the page to back out to the configuration screen. Click the **DONE** button at the bottom of the page to back out to the configuration screen.
 +----
 +=== 10. ===
 +In the device configuration screen, click the back arrow at the top left of the page to return to the device selection screen.
 </WRAP> </WRAP>
 <WRAP column half> <WRAP column half>
-{{ :reference:instrumentation:openlogger:initial_instrument_panel.png?500 |Initial Instrument Panel}}+{{ reference:test-and-measurement:openlogger:calibration_complete.png?500 |Calibration complete}}
 </WRAP> </WRAP>
 </WRAP> </WRAP>
  
-<WRAP group> +---- 
-<WRAP column half> +===== More Information ===== 
-=== 9. === + 
-In the device configuration screen, click the back arrow at the top left of the page to return to the device selection screen. +Check out the OpenLogger's [[test-and-measurement:openlogger:start|Resource Center]] for more guides, tutorials, and projects.
-</WRAP> +
-</WRAP>+
  
 +{{tag>openlogger}}
 {{include_n>4}} {{include_n>4}}