 ====== OpenLogger - Calibration ====== ====== OpenLogger - Calibration ======
  
 +===== First-Time Calibration =====
 +
 +Before being used for the first time, a new OpenLogger must first be calibrated. This process determines constants used to accurately calculate analog voltages for data points being logged, and stores these constants in non-volatile memory.
 +
 +**Note:** //After first-time calibration, described in this guide, a device can be recalibrated, if desired, through its Configuration Menu, found through the Device Manager in WaveForms Live.//
 +
 +----
 +
 +<WRAP group>
 +<WRAP column half>
 +=== 1. ===
 +When the device is connected to WaveForms Live for the first time, a prompt to calibrate the OpenLogger will appear. Click **OK**.
 +
 +</WRAP>
 +<WRAP column half>
 +{{ reference:test-and-measurement:openlogger:uncalibrated_device_prompt.png?500 |Calibrate your device}}
 +</WRAP>
 +</WRAP>
 +
 +<WRAP group>
 +<WRAP column half>
 +=== 2. ===
 +Connect the provided 2x18 flywires to the OpenLogger with the **__nub side facing towards the center__** of the OpenLogger.
 +----
 +=== 3. ===
 +Connect the red wire on pin 6 (DC OUTPUT 1) to the solid orange wire on pin 35 (Analog Input 1). Take a look at the [[test-and-measurement:openlogger:reference-manual#openlogger_pinout|OpenLogger Pinout]] in the Reference Manual for details on which wire corresponds to which OpenLogger pin.
 +</WRAP>
 +<WRAP column half>
 +{{ /learn/instrumentation/tutorials/openlogger/calib-1.png?500 |Connect the red and orange wires}}
 +
 +</WRAP>
 +</WRAP>
 +
 +<WRAP group>
 +<WRAP column half>
 +=== 4. ===
 +In WaveForms Live, click the **Begin** button to begin calibrating Analog Input 1.
 +</WRAP>
 +<WRAP column half>
 +{{ reference:test-and-measurement:openlogger:calibration_step_1.png?500 |Begin the calibration process}}
 +</WRAP>
 +</WRAP>
 +
 +<WRAP group>
 +<WRAP column half>
 +=== 5. ===
 +Wait for the calibration initialization to complete.
 +</WRAP>
 +<WRAP column half>
 +{{ reference:test-and-measurement:openlogger:starting_calibration_process.png?500 |Calibration progress bar}}
 +</WRAP>
 +</WRAP>
 +
 +<WRAP group>
 +<WRAP column half>
 +=== 6. ===
 +Connect the solid blue pin 33 (Analog Input 2 to the same red wire previously used and click the **Next** button.
 +
 +=== 7. ===
 +After calibrating Analog Input 2, continue through the remaining 6 analog inputs (the pink, green, gray, purple, yellow, and brown wires) until calibration is complete. Once all analog inputs have been calibrated, click the **DONE** button.
 +</WRAP>
 +<WRAP column half>
 +{{ /learn/instrumentation/tutorials/openlogger/calib-2.png?500 |Start step 2}}
 +</WRAP>
 +</WRAP>
 +
 +<WRAP group>
 +<WRAP column half>
 +=== 8. ===
 +A final calibration complete screen will appear. Click the **DONE** button - see the image to the right.
 +----
 +=== 9. ===
 +Click the **DONE** button at the bottom of the page to back out to the configuration screen.
 +----
 +=== 10. ===
 +In the device configuration screen, click the back arrow at the top left of the page to return to the device selection screen.
 +</WRAP>
 +<WRAP column half>
 +{{ reference:test-and-measurement:openlogger:calibration_complete.png?500 |Calibration complete}}
 +</WRAP>
 +</WRAP>
 +
 +----
 +===== More Information =====
 +
 +Check out the OpenLogger's [[test-and-measurement:openlogger:start|Resource Center]] for more guides, tutorials, and projects.
 +
 +{{tag>openlogger}}