-====== How to use a USB-TEMP, with a thermocouple and RTD sensor, to measure temperature with InstaCal and DAQami ======+====== How to Use a USB-TEMP, With a Thermocouple and RTD Sensor, to Measure Temperature With InstaCal and DAQami ======
  
-This article explains how to use a MCC [[http://www.mccdaq.com/usb-data-acquisition/USB-Temp.aspx | USB-TEMP]] module with a J-type thermocouple wire (MCC part # 745690-J001) and a 3-wire, 100 Ohm RTD sensor (MCC part # 745691-02).  For software, the latest release of the MCC DAQ CD was used to install InstaCal and DAQami, onto a Windows 10 system.  +This article explains how to use a MCC USB-TEMP module with a J-type thermocouple wire (MCC part # 745690-J001) and a 3-wire, 100 Ohm RTD sensor (MCC part # 745691-02).  For software, the latest release of the MCC DAQ CD was used to install InstaCal and DAQami, onto a Windows 10 system.  
  
 When installing the MCC DAQ CD (mccdaq.exe), select DAQami and the InstaCal & Universal Library packages. When installing the MCC DAQ CD (mccdaq.exe), select DAQami and the InstaCal & Universal Library packages.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_3a.png?446x420}}+{{https://files.digilent.com/reference-images/img_3a.png?446x420}}
  
 On a system with a 32-bit version of Windows OS, Measurement Computing’s software will appear in the folder path: ‘C:\Program Files\Measurement Computing’. On a system with a 32-bit version of Windows OS, Measurement Computing’s software will appear in the folder path: ‘C:\Program Files\Measurement Computing’.
Line 15: Line 15:
  
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_16.png?383x114}}+{{https://files.digilent.com/reference-images/img_16.png?383x114}}
  
 The next step is to launch MCC’s InstaCal software. The next step is to launch MCC’s InstaCal software.
Line 23: Line 23:
 If connecting a USB-TEMP module to the target system, for the first time, then you will be prompted to run the HID Registry Update utility. If connecting a USB-TEMP module to the target system, for the first time, then you will be prompted to run the HID Registry Update utility.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_18.png?479x370}}\\  +{{https://files.digilent.com/reference-images/img_18.png?479x370}}\\  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_19.png?632x327}}\\  +{{https://files.digilent.com/reference-images/img_19.png?632x327}}\\  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_20.png?391x295}}\\  +{{https://files.digilent.com/reference-images/img_20.png?391x295}}\\  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_21.png?391x295}}\\ +{{https://files.digilent.com/reference-images/img_21.png?391x295}}\\ 
  
 After rebooting the target system or cycling the USB cable connection to the module, restart InstaCal.  A ‘Plug and Play Board Detection’ window will appear.  Select the USB-TEMP device and click the OK button. After rebooting the target system or cycling the USB cable connection to the module, restart InstaCal.  A ‘Plug and Play Board Detection’ window will appear.  Select the USB-TEMP device and click the OK button.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_22.png?302x276}}+{{https://files.digilent.com/reference-images/img_22.png?302x276}}
  
 Once in InstaCal, the USB-TEMP module will appear in the PC Board List. Once in InstaCal, the USB-TEMP module will appear in the PC Board List.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_23.png?481x221}}+{{https://files.digilent.com/reference-images/img_23.png?481x221}}
  
 Right mouse-click on the USB-TEMP device to configure the device.   Right mouse-click on the USB-TEMP device to configure the device.  
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_24.png?480x408}}+{{https://files.digilent.com/reference-images/img_24.png?480x408}}
  
 In this example, channel 0 was configured to use a single 3-wire, 100 Ohm RTD sensor (MCC part # 745691-02).  Set the coefficient standard to ‘DIN 43760’ and click the ‘Show Connections’ button to view the appropriate wiring for the RTD sensor. In this example, channel 0 was configured to use a single 3-wire, 100 Ohm RTD sensor (MCC part # 745691-02).  Set the coefficient standard to ‘DIN 43760’ and click the ‘Show Connections’ button to view the appropriate wiring for the RTD sensor.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_25.png?691x463}}+{{https://files.digilent.com/reference-images/img_25.png?691x463}}
  
 Next, click on the CH(2-3) tab to configure channel 2 with a Type J thermocouple.   Next, click on the CH(2-3) tab to configure channel 2 with a Type J thermocouple.  
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_26.png?685x463}}+{{https://files.digilent.com/reference-images/img_26.png?685x463}}
  
 Click the OK button to exit the Board Configuration window.  A message window will prompt the user to save the configuration.  Click 'Yes'. Click the OK button to exit the Board Configuration window.  A message window will prompt the user to save the configuration.  Click 'Yes'.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_27.png?412x206}}+{{https://files.digilent.com/reference-images/img_27.png?412x206}}
  
 Using InstaCal’s top menu, select TEST – ANALOG. Using InstaCal’s top menu, select TEST – ANALOG.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_28.png?297x190}}+{{https://files.digilent.com/reference-images/img_28.png?297x190}}
  
 In the Board Test window, channel 0 was selected for the 3-wire RTD sensor and channel 2 was selected for the Type J thermocouple.  Click the TEST button to acquire temperature readings.  Click the OK button to exit the test window. In the Board Test window, channel 0 was selected for the 3-wire RTD sensor and channel 2 was selected for the Type J thermocouple.  Click the TEST button to acquire temperature readings.  Click the OK button to exit the test window.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_29.png?457x431}}+{{https://files.digilent.com/reference-images/img_29.png?457x431}}
  
 Close InstaCal and launch DAQami.  If this is the first time using DAQami, then a message window will appear with several option.  Click the OK button to continue in evaluation mode. Close InstaCal and launch DAQami.  If this is the first time using DAQami, then a message window will appear with several option.  Click the OK button to continue in evaluation mode.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_30.png?594x368}}+{{https://files.digilent.com/reference-images/img_30.png?594x368}}
  
 On the Getting Started window, click on the ‘New Configuration’ option followed by ‘Manual’. On the Getting Started window, click on the ‘New Configuration’ option followed by ‘Manual’.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_31.png?787x580}}+{{https://files.digilent.com/reference-images/img_31.png?787x580}}
  
 In the upper left-hand corner, click on the leftmost icon (Available Devices).  Click on the USB-TEMP device and drag the device onto the Device Panel, beneath the list. In the upper left-hand corner, click on the leftmost icon (Available Devices).  Click on the USB-TEMP device and drag the device onto the Device Panel, beneath the list.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_32.png?375x302}} +{{https://files.digilent.com/reference-images/img_32.png?375x302}} 
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_33.png?782x289}}+{{https://files.digilent.com/reference-images/img_33.png?782x289}}
  
 On the right-side, click on the ‘Channels’ tab.  Enable Analog Input channels ‘AI0’ and ‘AI2’ by selecting the Active checkbox.  For both channel 0 (the 3-wire RTD sensor) and channel 2 (the Type J thermocouple), set the Channel Settings TC Type to ‘J’.  On the right-side, click on the ‘Channels’ tab.  Enable Analog Input channels ‘AI0’ and ‘AI2’ by selecting the Active checkbox.  For both channel 0 (the 3-wire RTD sensor) and channel 2 (the Type J thermocouple), set the Channel Settings TC Type to ‘J’. 
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_34.png?621x445}}+{{https://files.digilent.com/reference-images/img_34.png?621x445}}
  
 Click on the ‘Samples Rate’ tab to view the rate range. Click on the ‘Samples Rate’ tab to view the rate range.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_35.png?587x234}}+{{https://files.digilent.com/reference-images/img_35.png?587x234}}
  
 In the upper left-hand corner, click on the second icon (Displays) from the left.   In the upper left-hand corner, click on the second icon (Displays) from the left.  
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_36.png?578x235}}+{{https://files.digilent.com/reference-images/img_36.png?578x235}}
  
 Next, click on the Scalar display option and drag it over to the Display Panel area, on the right.  Next, click on the Scalar display option and drag it over to the Display Panel area, on the right. 
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_37.png?616x377}}+{{https://files.digilent.com/reference-images/img_37.png?616x377}}
  
 Select an Active Channel and drag it onto the Scalar display. Select an Active Channel and drag it onto the Scalar display.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_38.png?711x349}}+{{https://files.digilent.com/reference-images/img_38.png?711x349}}
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_39.png?489x261}}+{{https://files.digilent.com/reference-images/img_39.png?489x261}}
  
 Above the Display Panel area, click the green arrow to ‘run the acquisition without logging to disk’. Above the Display Panel area, click the green arrow to ‘run the acquisition without logging to disk’.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_40.png?816x242}}+{{https://files.digilent.com/reference-images/img_40.png?816x242}}
  
 Otherwise, click the red circle to ‘run the acquisition and log the data to disk’. Otherwise, click the red circle to ‘run the acquisition and log the data to disk’.
  
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_41.png?777x233}} +{{https://files.digilent.com/reference-images/img_41.png?777x233}} 
-{{https://files.digilent.com/resources/measurement-computing/Reference Images/img_42.png?649x232}}+{{https://files.digilent.com/reference-images/img_42.png?649x232}}