Starting with the Electronics Explorer Board
1. Hardware
To set up the board:
First install Digilent WaveForms software on your PC. See
Installer Details for more information.
Connect a 12V 60W power supply to the Electronics Explorer board.
Connect the board to your PC using a USB cable.
The Ready
LED will light when the Output Enable switch is ON and the WaveForms application is connected to the board.
All of the components of the board work relative to the common ground. None of the instruments have a floating ground.
3. Troubleshooting
In case you receive the “Communication with device failed” error message:
Remove all of the attached components and wires from the device.
Verify that the USB cable is attached correctly, reconnect it.
Verify that the power supply is 12V at least 60W.
Connect with WaveForms to the device.
The board switch in the OFF position disables all the output. In this case, an “OFF” text is shown in the main window status bar and in the main plot of instruments.
4. Oscilloscope
The Oscilloscope has four channels with AC and DC input.
Specifications:
The input bandwidth is 100
MHz and the
ADC frequency is 40 MSps.
The DC inputs have an impedance of 9.3 MOhms in parallel with 10 pF.
The AC inputs have 100 nF filter capacitor.
The input voltage is from -20 V to 20 V.
The
ADC is 40
MHz, 10 bits with adjustable resolution from 800 uV to 40 mV.
The offset is adjustable from -20 V to 20 V.
The maximum input voltage is 200 V.
The buffer can be up to 16 kS (default is 8kS) on four channels.
The Electronics Explorer board is equipped with two Waveform Generators.
Specifications:
The output current is 25 mA.
The
DAC is 14 bits and 40 MSps.
The output 20Vp-p, 1.2 mV resolution.
The output bandwidth is 20
MHz.
When a channel is closed, the output is not in high impedance but should be close to zero volts. Verify the voltage on the Waveform Generator channels before connecting it to circuit.
The carrier buffer can be up to 32 kS (default is 8kS) or 16kS carrier and 16kS modulation buffer on two channels.
The positive and negative power supplies can be used as slow voltage or current basic waveform generators.
6. Power Supplies and Voltmeters
The board has four voltmeters, two reference voltages, one positive, one negative and one 3.3/5V supply.
Specifications:
The fixed power supply 3.3/5 V can provide up to 2 A.
The positive power supply has a range of 0 to 9 V and 0 to 1.5 A current limit.
The negative power supply has a range of 0 to -9 V and 0 to -1.5 A current limit.
The reference voltage supplies can provide only limited current, up to 10 mA.
The disabled reference voltage output is not in high impedance but should be close to zero volts.
7. Digital I/O
The Electronic Explorer board has 32 digital I/O (DIO 0-31) and 4 trigger I/O pins.
Specifications:
The digital pins are supplied at 3.3 V.
They have a 16 mA drive strength.
They have a 220 Ohm series thermistors.
They can output push-pull (PP), open-drain (OD), open-source (
OS), and three-state (TS) signals. For OD and
OS signals, use external pull-up or pull-down resistors.
The generator/analyzer frequency is 100
MHz.
The analyzer can work on an external clock, up to 100
MHz.
The Logic Analyzer buffer can be up to 16 kS (default is 4 kS).
The Pattern Generator custom buffer can be up to 16 kS (default is 1 kS).