These specifications are valid following 30 minutes of warm-up and are typical at 25 °C unless otherwise noted.
Supports the Oscilloscope, Voltmeter, Data Logger, Spectrum Analyzer, Network Analyzer, Impedance Analyzer, and Script Editor instruments.
| Number of Channels | ADP3450: 4 ADP3250: 2 There are no other specification differences between the ADP3450 and ADP3250. |
|
|---|---|---|
| Input Type | Single-ended | |
| Connector Type | Front panel BNC | |
| Input Coupling | AC, DC | |
| Bandwidth (-3dB) | ≥55 MHz1) | |
| Hardware Bandwidth Limit (-3dB) | 20 MHz, can be enabled or disabled in software per channel | |
| Resolution | 14 bits | |
| Software-Enhanced Vertical Resolution | 16 bits at ≤25 MS/s (at default 100 MS/s sample rate) | |
| Input Impedance | 1 MΩ ││ 15 pF | |
| AC Coupling Cut-off Frequency | 12 Hz at -3 dB | |
| Absolute Resolution | 0.125 mV (scale ≤ 200 mV/div) 3.05 mV (scale > 200 mV/div) |
|
| Accuracy (DC to 1 kHz, Integrative measurements)2) | ±10 mV ± 0.5% (scale ≤ 200 mV/div, VinCM = 0 V) ±100 mV ± 0.5% (scale > 200 mV/div, VinCM = 0 V) |
|
| Input Range | ±25 V with respect to ground (50 V peak-to-peak) | |
| Vertical Sensitivity (range) | 500 µV/div to 5 V/div (10 ranges) | |
| Acquisition Modes | Sample, peak detect, averaging, oversampling3) | |
| Overvoltage Protection | ±50 V DC or ±30 V RMS | |
| Range | Full Scale | Offset | Offset Accuracy |
|---|---|---|---|
| Low range (≤200 mV/div) | 2 V peak-to-peak | ±1 V | ±25 mV |
| High range (>200 mV/div) | 50 V peak-to-peak | ±25 V | ±625 mV |
| Default | Maximum | |
|---|---|---|
| Sample Rate (real-time sampling) | 100 MS/s per channel in Repeated/Shift/Screen Modes | 125 MS/s per channel in Repeated/Shift/Screen Modes |
| 100 MS/s total in Record Mode 1) | 125 MS/s total in Record Mode 2) | |
| Sample Rate (equivalent-time sampling - oversampling) | 400 MS/s per channel | 0.5 GS/s per channel |
| Buffer Size | 32,768 samples per channel in Repeated/Shift/Screen Modes |
|---|---|
| 128 MS 3) total in Record Mode |
Note: Acquisitions are made simultaneously on all enabled channels in all modes.
Supports the Logic Analyzer, Pattern Generator, Static I/O, Protocol Analyzer, and Script Editor instruments.
| Number of Channels | 16 |
|---|---|
| Connector | Header 100 mil 2×12 MTE |
| Input Voltage | 0 V to 3.3 V (5 V tolerant) |
| Input Current | ≤50 µA |
| Input Threshold | 0 V to 2.0 V, dependent on adjustable supply rail |
| Input Type | LVCMOS (1.2 V to 3.3 V, 5 V tolerant), dependent on adjustable supply rail |
| Input Logic Level | Input Low Voltage, VIL Min 0 V Max 0.8 V Input High Voltage, VIH Min 2.0 V Max 5.25 V |
| Output Type | Adjustable 1.2 V to 3.3 V set through adjustable supply rail LVCMOS (3.3 V, 8 mA) |
| Output Logic Level | Output Low Voltage, VOL sinking 8 mA Min 0 V Max 0.4 V Output Max Voltage, VOH sourcing 8 mA Min 2.4 V Max 3.456 V |
| DIO Channel Pull Resistors | 10 kΩ, pull-up to adjustable supply rail |
| Function Control | Individually programmable as Digital I/O, Logic Analyzer or Pattern Generator |
| Protocols | SPI, I2C, UART, CAN, I2S, 1-Wire, HDMI CEC, Manchester codes, custom 1) |
| Pattern Generator | Constant, clock, pulse, random, number, Binary counter, Gray counter, Johnson counter, Decimal counter, walking 0/1, custom |
| Custom Patterns File | Import and export custom data as *.csv, *.txt or *.tdms file |
| Direction Control | Input or output, individually selectable |
| Channel-to-Channel Skew | 2 ns, typical |
| Overvoltage Protection | Short-circuit to ground, ±15 V |
Note: Characteristics of the logic standard used for digital input/output can be configured within the Power Supplies instrument. For the digital power supply's specifications, see the Digital Power Supply section, below.
| Default | Maximum | |
|---|---|---|
| Sampling Rate (real-time sampling) | 100 MS/s per channel | 125 MS/s per channel |
| Logic Analyzer Buffer Memory | 32,768 samples per channel in Repeated/Shift/Screen Modes |
|---|---|
| 64 MS 1) total in Record Mode |
| Pattern Generator Buffer Memory | 16,384 samples per channel |
|---|
| Trigger Sources | Oscilloscope analog channels, Function generator start, Digital I/O lines, External triggers (TRIG1/TRIG2), Manual |
|---|---|
| Trigger Modes | None, Auto, Manual (Forced Trigger), Single |
| Analog Trigger | Edge, pulse, transition, condition, level, hysteresis, hold-off |
| Digital Trigger | Edge, level, pattern, glitch |
| Analog/Oscilloscope Trigger Resolution | 2 ns |
| Digital/Logic Analyzer Trigger Resolution | 8 ns |
| Number of Channels | 2 |
|---|---|
| Trigger Type | Digital |
| External Trigger Connector | Rear panel BNC |
| Input Voltage | 3.3 V, 5 V tolerant |
| Input Current | ≤50 µA |
| Input Threshold | 0 V to 2.0 V |
| Input Type | LVCMOS (3.3 V, 5 V tolerant) |
| Input Logic Level | Input low voltage, VIL Min 0 V Max 0.8 V Input high voltage, VIH Min 2.0 V Max 5.25 V |
| Output Type | LVCMOS (3.3 V, 8 mA) |
| Output Logic Level | Output low voltage, VOL Min 0 V Max 0.4 V Output high voltage, VOH Min 2.4 V Max 3.465 V |
| Channel Pull Resistors | 1 MΩ pull-down |
| Direction Control | Individually programmable as input or output |
| Overvoltage Protection | Short-circuit to ground, ±15 V |
Supports the Waveform Generator, Network Analyzer, Impedance Analyzer, and Script Editor instruments.
| Number of Channels | 2 |
|---|---|
| Output Type | Single-ended |
| Connector Type | Front panel BNC |
| Update Rate | 100 MS/s by default, 125 MS/s maximum |
| Resolution | 14 bits |
| Output Impedance | 50 Ω |
| Bandwidth (-3 dB) | >15 MHz |
| Standard Functions | Sine, square, triangle, sawtooth, ramp up, ramp down, DC voltage, noise, trapezium, others |
| Advanced Waveforms | Sweep, modulation (AM, FM), math, play mode, custom |
| Sweep Modes | Frequency and Amplitude. Up and down with selectable start/stop frequencies and settable time increments |
| Custom Waveform Files Supported | Import files *.csv, *.txt, *.mp3, *.wav, *.wmv & *.avi, export as image, or as raw data in *.csv, *.txt or *.tdms formats |
| Output Frequency Resolution | 1 µHz |
| Output Voltage Range | ±5 V with open circuit or high-Z load. ±3.5 V at maximum current. |
| DC Offset Range | ±5 V |
| DC Current Drive | 30 mA maximum |
| Slew Rate | 200 V/us (5 V step) |
| Absolute Resolution | 150 µV (│Vout│ ≤ 1 V) 670 µV (│Vout│ > 1 V) |
| Accuracy | ±10 mV ± 0.5% (│Vout│ ≤ 1 V) ±25 mV ± 0.5% (│Vout│ > 1 V) |
| Buffer Size | 32,768 samples per channel |
| Power-Off State | High Impedance |
| Overvoltage Protection | Short-circuit to ground, ±15 V |
| Range Setting | AC Amplitude Range | DC Offset Range | Resolution | Amplitude Error | DC Offset Error | Total Output Range |
|---|---|---|---|---|---|---|
| High range | ±5 V | ±5 V | 670 µV/LSB | ±0.5% | ±50 mV | ±5 V |
| Low range | ±1 V | ±5 V | 150 µV/LSB | ±0.5% | ±10 mV | ±5 V |
Note: Shares external triggers with Mixed Signal Oscilloscope.
Shares digital input/output channels with Mixed Signal Oscilloscope: See the Digital Channels specifications.
| Frequency Range | 0 Hz to 50 MHz |
|---|---|
| Display Modes | Magnitude, average, peak hold, min hold, count |
| Y Axis | Logarithmic (dBṼ, dBu, dBm) or linear (volts) |
| X Axis | Linear or Logarithmic |
| Windowing Functions | Rectangular, Triangular, Hamming, Hann, Cosine, Blackman-Harris, Flat Top, Kaiser |
| Frequency Range | 10 µHz to 10 MHz, up to 10,001 samples |
|---|---|
| Display Modes | Magnitude, Phase |
| Y Axis | Linear or Logarithmic |
| X Axis | Linear or Logarithmic |
| Plots | Time, FFT, Nichols, Nyquist |
Shares digital input/output channels with Mixed Signal Oscilloscope: See the Digital Channels specifications.
| Protocols | UART, SPI, I2C, CAN, AVR |
|---|
| Operations | Addition “+”, Subtraction “-”, Multiplication “*”, Division “/”, Remainder “%” |
|---|---|
| Brackets | Parenthesis “()”, Square “[]” |
| Constants | Exp, Ln, Log, Pi |
| Functions | Logarithm, power, minimum, maximum, square root, sine, cos, tan, arccos, arctan, arctan2, absolute value, round, floor, ceil |
| Operands | All analog and digital input channels, reference waveforms, time, constants, Pi |
| Oscilloscope Mode | DC to maximum bandwidth of scope |
|---|---|
| Spectrum Mode | Magnitude, average, peak hold, min hold, count |
| Statistics | Logarithmic (dBṼ, dBu, dBm) or linear (volts) |
Supported by the Power Supplies and Script Editor instruments. Contains logic standard settings for the digital input/output channels. See the Digital Channels specifications.
| Number of Channels | 1 |
|---|---|
| Voltage Range | 1.2 V to 3.3 V |
| Current Output | 300 mA |
| Connector Type | 2 pins included in the Digital 100 mil 2×12 MTE Header |
| Device to Computer Interface | USB 2.0 Hi-Speed, Type B |
|---|---|
| USB 2.0 Host Ports | 4 USB 2.0 Hi-Speed Type A (*see below) |
* USB Host Port Current Limits
| Minimum | Typical | Maximum | |
|---|---|---|---|
| Left Ports (Indicated by circle on enclosure) | 0.9 A | 1.06 A | 1.22 A |
| Right Ports | 0.51 A | 0.61 A | 0.68 A |
Note: Digilent recommends the use of the bottom left USB host port for connection to Wi-Fi dongles.
| Network Interface | 1000 Base-TX, full-duplex 100 Base-TX, full-duplex 100 Base-TX, half-duplex 10 Base-T, full-duplex 10 Base-T, half-duplex |
|---|---|
| Communication Rates | 10/100/1000 Mbps, auto-negotiated |
| Connector Type | RJ-45 |
Note: MAC address located on a sticker next to the Ethernet port.
The ADP3450 and ADP3250 require an auxiliary power supply.
| Power Supply Voltage | 19 V |
|---|---|
| Power Supply Current | 3.4 A |
| Power Consumption | 45 W maximum |
| Barrel Connector Size | 5.5 mm x 1.65 mm |
| Dimensions | 23.40 cm x 19.40 cm x 3.81 cm (9.2 in x 7.6 in x 1.5 in) |
|---|---|
| Weight | 450 g |
| Operating Temperature | 0 °C to 40 °C (32 °F to 104 °F) |
|---|---|
| Storage Temperature | -20 °C to 60 °C (-4 °F to 140 °F) |
| Operating Humidity | 10% to 80% RH non-condensing |
| Storage Humidity | 5% to 95% RH non-condensing |
| Cooling | Convection-cooled. Ensure the ventilation holes are not obstructed. 1) 2) |
| Pollution Degree | 2 |
| Maximum Altitude | 2,000 m |
Indoor use only
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
This product meets the essential requirements of applicable European Directives, as follows: